Verify correct detector for spectral range, scan speed, is selected in
software. Slow response detectors, such as DTGS detectors, will not response
well at high scan speeds.
Verify detector is positioned correctly and connected. Check power at
detector.
Follow infrared beam (conceptually, as it is not visible). Make sure path is
unobstructed. Does problem exist with accessories removed?
Verify source is on. If not on, remove and test for open source element. If
source is OK, suspect power supply is bad.
Inspect beamsplitter for fogging, replace if necessary.
Cool liquid nitrogen detector.
Swap detector/electronics/power supplies, one at a time.